Overview

A factor that is used to calculate proceeding dilutions during olfactometry analysis.  A ratio of the current dilution to the step factor calculates the next dilution in the dilution series.

Example of Step Factors

Max dilution = 400
Min dilution=53
Step Factor = 1.5
Steps or Dilution Series = 6

Dilution4002671781187953
Step Factor 1.5 400/1.5 = 267267/1.5 = 178178/1.5 =118118/1.5 = 7979/1.5 = 53
Step/Dilution Series =6123456